Instruction Manual
012-09384D
Water Quality MultiMeasure Sensor
PS-2169
WATER
QUALITY
PS-2169
ä
SENSOR
CONDUCTIVITYpH/ISE/ORP
TEM
PERATURE
DISSOLVED OXYGEN
Included Equipment Part Number
Water Quality MultiMeasure Sensor PS-2169
Stainless Steel Temperature Probe PS-2153
pH Electrode 699-195
Dissolved Oxygen (DO
2
) Probe
1
1
The DO
2
Probe comes with a bottle of filling solution, a syringe, a replacement membrane cartridge
and O-ring, and a plunger. See illustration on page 7.
699-06320
Conductivity Electrode (10x) 699-06621
Required Equipment
PASPORT Interface
2
2
PASPORT interfaces include the AirLink (IW-4100 or PS-2005), Xplorer GLX (PS-2002), Xplorer
(PS-2000), PowerLink (PS-2001), and USB Link (PS-2100)
See PASCO catalog or
www.pasco.com
Optional Equipment
Fast-response Temperature Probe PS-2135 (3-pack)
Skin/surface Temperature Probe PS-2131
Conductivity Electrode (1x) 699-06620
Various Ion-selective Electrodes (ISE) See PASCO catalog or
www.pasco.com
Oxidation Reduction Potential (ORP) Electrode CI-6716
DO
2
Electrode Maintenance Kit CI-6541
Zobrazit stránku 0
1 2 3 4 5 6 ... 20 21

Shrnutí obsahu

Strany 1 - Instruction Manual

®Instruction Manual012-09384DWater Quality MultiMeasure SensorPS-2169WATERQUALITYPS-2169äSENSORCONDUCTIVITYpH/ISE/ORPTEMPERATUREDISSOLVED OXYGENInclud

Strany 2 - Introduction

®Water Quality MultiMeasure Sensor DataStudio Calibration10Appendix A: DataStudio CalibrationIn DataStudio, click the Setup button to open the Experim

Strany 3 - Conductivity

®Model No. PS-2169 EZ Screen11One-point CalibrationA one-point slope calibration requires one known standard. For conductivity, use one of the soluti

Strany 4

®Water Quality MultiMeasure Sensor Reference Solutions for Conductivity Calibration12Appendix C: Reference Solutions for Conductivity CalibrationAppen

Strany 5 - Measuring pH

®Model No. PS-2169 DO2 Calibration Tables135.0 13.3 13.3 13.2 13.1 13.0 12.9 12.8 12.7 12.7 12.6 12.5 12.4 12.3 12.2 12.2 12.1 12.0 11.9 11.8 11.75.5

Strany 6 - Dissolved Oxygen

®Water Quality MultiMeasure Sensor DO2 Calibration Tables1427.0 8.38.38.28.28.18.08.07.97.97.87.87.77.77.67.67.57.57.47.37.327.5 8.28.28.18.18.08.07.9

Strany 7

®Model No. PS-2169 DO2 Calibration Tables155.5 11.5 11.4 11.3 11.2 11.2 11.1 11.0 10.9 10.8 10.7 10.7 10.6 10.5 10.4 10.3 10.2 10.2 10.1 10.0 9.96.0

Strany 8

®Water Quality MultiMeasure Sensor DO2 Calibration Tables16Table 2: DO2 Salinity Correction Factors 27.0 7.2 7.2 7.1 7.1 7.0 7.0 6.9 6.9 6.8 6.7 6.7 6

Strany 9 - Specifications

®Model No. PS-2169 DO2 Calibration Tables1710.0 1.000 0.996 0.993 0.989 0.986 0.982 0.979 0.975 0.971 0.968 0.964 0.960 0.957 0.953 0.949 0.946 0.942

Strany 10 - Two-point Calibration

®Water Quality MultiMeasure Sensor DO2 Calibration Tables189.0 0.938 0.934 0.930 0.926 0.923 0.919 0.915 0.911 0.907 0.904 0.900 0.896 0.892 0.888 0.8

Strany 11 - Appendix B: EZ Screen

®Model No. PS-2169 DO2 Calibration Tables197.0 0.871 0.867 0.863 0.859 0.855 0.851 0.847 0.843 0.839 0.835 0.831 0.828 0.824 0.820 0.816 0.812 0.8088

Strany 12 - Calibration Tables

®Water Quality MultiMeasure Sensor Introduction2IntroductionThe PS-2169 Water Quality MultiMeasure Sensor combines four sensors in a single unit:•Temp

Strany 13 - Model No. PS-2169 DO

®Water Quality MultiMeasure Sensor DO2 Calibration Tables205.0 0.801 0.797 0.793 0.789 0.785 0.781 0.777 0.773 0.769 0.765 0.761 0.757 0.753 0.749 0.7

Strany 14

®Model No. PS-2169 Technical Support21Technical SupportFor assistance with any PASCO product, contact PASCO at:Limited WarrantyFor a description of t

Strany 15

®Model No. PS-2169 Temperature3Xplorer GLX SetupIf you are using an Xplorer GLX in standalone mode, press to start data collec-tion. Note that the

Strany 16

®Water Quality MultiMeasure Sensor pH/ISE/ORP4the end of the electrode in the solution to be measured and wait for the reading to sta-bilize. Rinse th

Strany 17

®Model No. PS-2169 pH/ISE/ORP5Measuring pH The pH electrode produces a voltage proportional to the pH of the solution that it is immersed in. This vo

Strany 18

®Water Quality MultiMeasure Sensor Dissolved Oxygen6Dissolved OxygenThe DO2 component of the multi-sensor is designed for use in aqueous media at tem-

Strany 19

®Model No. PS-2169 Dissolved Oxygen7DO2 CalibrationPerform a one-point calibration using 100% humidified air as the known standard, which is equivale

Strany 20

®Water Quality MultiMeasure Sensor Dissolved Oxygen82. Rinse the electrode with tap water (or deionized water in areas with hard tap water) and rub it

Strany 21 - Technical Support

®Model No. PS-2169 Specifications9SpecificationsGeneralComponent Sensors Temperature, Conductivity, pH/ISE/ORP probe voltage, Dissolved OxygenMax. Sa